VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY
|
Back to Search Results |
|
Model Number 3100A |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/22/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
"vyaire file identification: (b)(6).The device was not be returned for further evaluation; therefore a definitive root cause could not be determined and corrective or preventive action is not indicated at this time.However, the customer checked the driver assy and discovered that the resistance is approximately 3 ohm.Checked the power supply board and discovered +5v, +15v, and -15v.Performed pressure calibration and piston position and displacement display calibration on r1 and r7 but not on r4.The customer then requested troubleshooting tips and a service manual for this new series of hfov as soon as possible.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
The customer reported to vyaire medical that the 3100a ventilator is not producing oscillation and that pressure is building up in the circuit.The circuit was calibrated, but when starting the oscillator, it does not start, and the "oscillator stopped" visual and audio alarms sound.Furthermore, there was no patient associated with the reported event.
|
|
Search Alerts/Recalls
|
|
|