Model Number B12LT
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2022
Event Type  malfunction  
Event Description
It was reported that during a laparoscopic appendectomy, the seal cap of the trocar was easy to loosen.No patient consequences.
 
Manufacturer Narrative
(b)(4).Date of event unknown, assumed first day of month that complaint was reported batch # unk.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
(b)(4).Date sent: 12/23/2022.D4: batch # unk.Investigation summary; the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample revealed that only the sleeve from the b12lt device was returned with the universal seal component disassembled inside a plastic bag and was observed to be not properly welded.In addition, the packaging opened was returned along with the instrument.No conclusion could be reached regarding what may have caused the reported incident.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key15923330
MDR Text Key307994707
Report Number3005075853-2022-08342
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier30705036001168
UDI-Public10705036001164
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberB12LT
Device Catalogue NumberB12LT
Device Lot Number902A68
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/09/2022
Initial Date FDA Received12/06/2022
Supplement Dates Manufacturer Received12/07/2022
Supplement Dates FDA Received12/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/03/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-