ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
|
Back to Search Results |
|
Model Number B12LT |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a laparoscopic appendectomy, the seal cap of the trocar was easy to loosen.No patient consequences.
|
|
Manufacturer Narrative
|
(b)(4).Date of event unknown, assumed first day of month that complaint was reported batch # unk.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 12/23/2022.D4: batch # unk.Investigation summary; the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample revealed that only the sleeve from the b12lt device was returned with the universal seal component disassembled inside a plastic bag and was observed to be not properly welded.In addition, the packaging opened was returned along with the instrument.No conclusion could be reached regarding what may have caused the reported incident.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.
|
|
Search Alerts/Recalls
|
|
|