It was reported to aesculap inc.That two (2) prestige atra grasper dbl-act 5mm (item # 8360-10) devices were used during a procedure performed on (b)(6) 2022.According to the complainant, during fibroid retrieval, the grasper jaws broke.The procedure was continued using a second prestige grasper, but the same issue recurred; jaw breakage observed.The complaint devices have been returned to the manufacturer for evaluation.No patient complications were reported as a result of this event.However, a surgical delay of ten (10) minutes was noted.The malfunction is filed under aic reference (b)(4).Reference associated medwatch report # 2916714-2022-00114.
|
Manufacturing site evaluation: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
|