MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD BUBBLE CPAP SYSTEM; BZD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD BUBBLE CPAP SYSTEM; BZD Back to Search Results
Model Number BC163
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/09/2022
Event Type  malfunction  
Event Description
A healthcare facility in colorado reported, via a fisher & paykel healthcare (f&p) field representative, that the cpap probe of two bubble cpap generator's was loose.The bubble cpap generator is part of the bc163 bubble cpap system kit.There was no patient involvement as the issue was found whilst the device was not in use on a patient.
 
Manufacturer Narrative
(b)(4).The complaint bc163 bubble cpap system kit is currently en route to fisher and paykel healthcare (f&p) for investigation.We will provide a follow up report upon completion of our investigation.
 
Manufacturer Narrative
(b)(4).Method: the complaint bc163-10 bubble cpap system was returned to fisher & paykel healthcare (f&p) in new zealand for evaluation, where it was visually inspected, and performance tested by a trained f&p technician.Results: the measured dimensions of the bc163-10 bubble cpap systems' generator was out of specification.Conclusion: we are unable to determine the cause of the reported fault.The existing design of the cpap probe is intended to be adjustable, which leads to the possibility of inadvertent movement.We note that there is a physical stop in the bubble cpap probe, to prevent the pressure from exceeding 10cmh20, and that a pressure manifold is used with the breathing circuit, to reduce the risk of unsafe circuit pressure.The user instructions that accompany the bubble cpap system kit illustrate in pictorial format the correct set-up and proper use of the bubble cpap generator.It also states the following: "always use pressure monitoring to verify that the patient is receiving the prescribed cpap level." "regularly observe the cpap generator for bubbling.If bubbling is not observed, check for and minimize the air leaks in the system and at the patient.If air leaks have been minimized, air flow may be increased to achieve continuous bubbling.".
 
Event Description
A healthcare facility in colorado reported, via a fisher & paykel healthcare (f&p) field representative, that the cpap probe of two bubble cpap generator's was loose.The bubble cpap generator is part of the bc163 bubble cpap system kit.There was no patient involvement as the issue was found whilst the device was not in use on a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BUBBLE CPAP SYSTEM
Type of Device
BZD
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534000
MDR Report Key15927104
MDR Text Key305069758
Report Number9611451-2022-01116
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBC163
Device Catalogue NumberBC163
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 11/09/2022
Initial Date FDA Received12/06/2022
Supplement Dates Manufacturer Received12/16/2022
Supplement Dates FDA Received12/20/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-