Model Number CV-180
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/09/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned as part of the asset return process.During routine inspection of the device by olympus, it was found that the leds (light emitting diodes) on the front panel were continuously on; this was due to a defective printed circuit board and defective front panel.A green display also appeared while shaking the cable of the power supply and required replacement.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The evis exera ii video system center device was returned as part of the asset return process.During routine inspection of the device by olympus, it was found that the leds (light emitting diodes) on the front panel were continuously on.There was no procedural or patient involvement associated with this event.This mdr is being submitted to capture the reportable malfunction found during the device evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the faulty front panel printed circuit board could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15927368
MDR Text Key308070117
Report Number3002808148-2022-04919
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170215513
UDI-Public04953170215513
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/09/2022
Initial Date FDA Received12/06/2022
Supplement Dates Manufacturer Received12/16/2022
Supplement Dates FDA Received01/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-