AVANOS MEDICAL INC. UNKNOWN CATHETERS Back to Search Results
Model Number UNKNOWN
Device Problem Material Integrity Problem (2978)
Patient Problem Insufficient Information (4580)
Event Date 11/06/2022
Event Type  malfunction  
Event Description
It was reported that on (b)(6) 2022, during the procedure, the surgeon placed two catheters bilaterally using tunnelers; the surgeon experienced difficulty placing one side.The catheters were removed on (b)(6) 2022 and the catheter which had been difficult to place broke during removal; there was no reported injury.Additional information received on (b)(6) 2022, the doctor reported, the patient was currently under palliative care and there was no plan to remove the tip of the catheter.The doctor additionally reported that he examined the catheter himself and it appeared that less than 5 mm was still possibly left inside if any was still left inside.The patient was due to be seen in the clinic in a week.
 
Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.Root cause could not be determined.All information reasonably known as of (b)(6) 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to (b)(4), in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.Device not returned.
 
Manufacturer Narrative
The product involved in the report has been returned and is being processed for evaluation.All information reasonably known as of (b)(6) 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Manufacturer Narrative
The actual sample from the reported event was returned for evaluation.Visual examination of the device revealed: the black tip was not present and the entire mid-body segment and infusion segment were stretched.The reported event could be confirmed as reported; broken tip; the root cause is undetermined.All information reasonably known as of 09 feb 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN CATHETERS
Type of Device
CATHETER
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S. DE R.L. DE C.V.
ave noruega edificio d-1b
fraccionamiento rubio
tijuana b.c.
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key15933672
MDR Text Key307835172
Report Number2026095-2022-00133
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/07/2022
Initial Date FDA Received12/07/2022
Supplement Dates Manufacturer Received12/21/2022
01/19/2023
Supplement Dates FDA Received01/05/2023
02/09/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age52 YR
Patient SexMale
Patient Weight80 KG
-
-