Model Number VTICM5_13.2
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problems Intraocular Pressure Increased (1937); Pupillary Block (2026); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 09/17/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).Claim# (b)(4).
 
Event Description
The reporter indicated that the surgeon implanted a 13.2mm vticm5_13.2; -14.0/1.5/105 (sphere/cylinder/axis) implantable collamer lens into the patient's left eye (os) on (b)(6) 2022.The patient experienced excessive vaulting; significant reduction of irido-corneal angles; pupil block with elevated intraocular pressure; fixed (unreactive) pupil; enlargement of pi - yag which was performed on (b)(6) 2022; and on (b)(6) 2022 the lens was explanted.On (b)(6) 2022 a replacement lens of shorter length lens was implanted but the problem was not resolved.The cause of the event was reported as unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene enriquez
1911 walker avenue
monrovia, CA 91016
MDR Report Key15936540
MDR Text Key305040383
Report Number2023826-2022-04499
Device Sequence Number1
Product Code QCB
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberVTICM5_13.2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/28/2022
Initial Date FDA Received12/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/04/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age38 YR
Patient SexMale
-
-