BOSTON SCIENTIFIC CORPORATION SPYSCOPE DS II ACCESS & DELIVERY CATHETER; CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID Back to Search Results
Model Number M00546610
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/11/2022
Event Type  malfunction  
Event Description
Note: this report pertains to two spyscope ds ii and a spyglass ds controller used during the same procedure.It was reported to boston scientific corporation that the two spyscope ds ii and a spyglass ds controller were used in the bile duct during a choledochoscopy procedure performed on (b)(6) 2022.During the procedure, when the spyscope ds ii was connected into the controller, image interference immediately appeared.A second spyscope ds ii was used; however, the problem continued and the image was not recovered.The procedure was not completed due to this event.There were no patient complications reported as result of this event.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPYSCOPE DS II ACCESS & DELIVERY CATHETER
Type of Device
CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15940848
MDR Text Key308238590
Report Number3005099803-2022-07061
Device Sequence Number1
Product Code FBN
UDI-Device Identifier08714729965404
UDI-Public08714729965404
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K183636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00546610
Device Catalogue Number4661
Device Lot Number0029936527
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/11/2022
Initial Date FDA Received12/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-