Reportable based on the device analysis completed on 15nov2022.It was reported that pump burst occurred.The target lesion was located in the radial anteriovenous fistula.An angiojet avx was used for the thrombectomy.However, during the procedure.The pump burst.The procedure was completed with another of the same device.There were no patient complications reported.However, returned device analysis revealed a hypotube break.
|
Lot number: reported as either 29754307, 29747046, or 29747047.Device evaluated by mfr: returned product consisted of an angiojet avx catheter.The pump assembly, effluent/supply line, shaft, tip, and spike line were visually examined for damage or any irregularities.The catheter shaft showed multiple bends and kinks.Functional testing was attempted; however, the device would not prime.During analysis at the severely kinked location at 38 cm from the tip the hypotube was clearly broken and completely separated.The device was x-rayed to verify the broken hypotube.The device could not be functionally tested due to the extreme damage on the device.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|