• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC DE COSTA RICA S.R.L. POLARMAP; CARDIAC MAPPING CATHETER, PERCUTANEOUS, SINGLE-USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC DE COSTA RICA S.R.L. POLARMAP; CARDIAC MAPPING CATHETER, PERCUTANEOUS, SINGLE-USE Back to Search Results
Lot Number 0029840174
Device Problems Material Perforation (2205); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2022
Event Type  malfunction  
Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Event Description
It was reported that during preparation for a cryo-ablation procedure to treat atrial fibrillation (a fib) using a polarmap catheter the shaft was kinked, the insulation was broken at the kinked portion.The catheter was replaced, prior to use inside the patient, and the procedure was completed with no patient complications.The catheter is not expected to be returned for analysis as it was disposed of by the site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARMAP
Type of Device
CARDIAC MAPPING CATHETER, PERCUTANEOUS, SINGLE-USE
Manufacturer (Section D)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
302 parkway
global park, la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key15942714
MDR Text Key307983304
Report Number2124215-2022-49855
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/27/2023
Device Lot Number0029840174
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2022
Initial Date FDA Received12/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-