RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).
|
|
Event Description
|
Philips received a complaint from the customer, reporting that the device had a loose gas outlet port.It was unknown if the device was in use at the time of the event.No patient or user harm reported.
|
|
Manufacturer Narrative
|
H10: there was no patient involvement when the issue occurred.No patient or user harm reported.The field service engineer (fse) reported performing annual preventative maintenance (pm) by replacing the front bezel, battery, and annual pm kit.During follow up, no further details were provided regarding the loose gas outlet port.The system meets the specification for the performed service and is returned to use.
|
|