RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Failure to Read Input Signal (1581)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/16/2022 |
Event Type
malfunction
|
Event Description
|
The system goes out of standby mode.Per remote service engineer: advised customer when onsite to check average air reading in pneumatics to see if in specs, can also run pvt tests 7,8,9 and suggested to replace flow sensor assembly if not passing and to correct issue.Investigation is ongoing.
|
|
Manufacturer Narrative
|
H10 multiple attempts have been made to try to obtain further information about this case but no response received from the customer.This file is closed and can be reopened if new information becomes available.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.H11: updated contact information, contact office entity, and manufacturing site.
|
|