• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRUMPF MEDIZIN SYSTEME GMBH + CO. KG TRULIGHT TRIBLE TUSM; SURGICAL LIGHT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRUMPF MEDIZIN SYSTEME GMBH + CO. KG TRULIGHT TRIBLE TUSM; SURGICAL LIGHT Back to Search Results
Model Number 1574851
Device Problem Flaked (1246)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2022
Event Type  malfunction  
Event Description
The customer alleged head chipping/peeling paint.No harm or impact to the surgical procedure was reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Manufacturer Narrative
The investigation identified steps in the preparation for coating process which can lead to lack of paint adhesion resulting in paint chipping.Additionally, collisions of the light heads and use of unapproved cleaners can further contribute to paint chipping from the light system.The damage to the painted surfaces usually does not develop suddenly but develops over time.The instructions for use state that the devices must be checked for proper condition prior to each use.Damaged devices must not be used.If damage to the surface coating is recognized and removed, there is no danger to the patient.The surface coating process has been updated to include additional testing and verification.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIGHT TRIBLE TUSM
Type of Device
SURGICAL LIGHT
Manufacturer (Section D)
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
carl zeiss strasse 7-9
saalfeld thuringen 07318
GM  07318
Manufacturer Contact
frances coote
carl zeiss strasse 7-9
saalfeld thuringen 07318
GM   07318
MDR Report Key15958581
MDR Text Key308148507
Report Number9681407-2022-00028
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1574851
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/07/2022
Initial Date FDA Received12/12/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-