AIZU OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V3
Device Problem Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2022
Event Type  malfunction  
Event Description
The customer reported to olympus, the uretero-reno videoscope experienced a leakage approximately 1 centimeter from the distal end.The event occurred during reprocessing.There was no report of patient harm associated with this event.During incoming inspection, it was determined the bending steel structure broke through bending section cover.This medwatch is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
The subject device was returned to olympus for evaluation.During inspection and testing, the allegation of leakage approximately 1 centimeter from the distal end was not confirmed.In addition to the finds reported in the event, the dial unit was damaged.The bending section cover was leaking.The universal cord was wrinkled.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was presumed to have been due to excessive force was applied to the passive bending section.Then the cable support got separated and protruded from the a-rubber (distal sheath).The user may be able to reduce / prevent occurrence of the suggested event by handling device in accordance with the following instructions for use (ifu): "·do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.·never perform angulation control forcibly or abruptly.Never forcefully pull, twist, or rotate the angulated bending section.Patient injury, bleeding, and/or perforation may result.It may also become impossible to straighten the bending section during an examination." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15959898
MDR Text Key305665035
Report Number9610595-2022-05252
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170403385
UDI-Public04953170403385
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/13/2022
Initial Date FDA Received12/12/2022
Supplement Dates Manufacturer Received01/13/2023
Supplement Dates FDA Received02/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/03/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-