Model Number 8103 |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/18/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
|
|
Event Description
|
It was reported that patient's device is believed to have prematurely depleted.The patient was seen to have really low battery life (ifi = yes) on (b)(6) 2022 visit.No known relevant surgical intervention has occurred to date.No other relevant information has been received to date.
|
|
Event Description
|
Tablet data was received and reviewed.A premature depletion event has been identified.No obvious cause has been established, but a report of shoulder surgery being used in tandem with an abnormal voltage threshold value being seen in the data; electrocautery use during this shoulder surgery is probable, and the likely cause of the premature end of a life.This will not be considered the root cause until further confirmation is received.No other relevant information has been received to date.
|
|
Event Description
|
After further review of data, it was determined that the battery life begins to rebound, indicating that a beginning of life impedance issue was the cause of the reported premature end of life.However, with this conclusion, it can be determined that no true premature end of life is occurring.This event is related to an increased duration of the high battery impedance experienced by cfx batteries during the beginning of life (bol) or an internal short that managed to ¿burn¿ itself out.This is not considered a fault of the device.No other relevant information has been received to date.
|
|