• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. SYMMETRY GEN MODEL 8103; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. SYMMETRY GEN MODEL 8103; GENERATOR Back to Search Results
Model Number 8103
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/18/2022
Event Type  malfunction  
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
It was reported that patient's device is believed to have prematurely depleted.The patient was seen to have really low battery life (ifi = yes) on (b)(6) 2022 visit.No known relevant surgical intervention has occurred to date.No other relevant information has been received to date.
 
Event Description
Tablet data was received and reviewed.A premature depletion event has been identified.No obvious cause has been established, but a report of shoulder surgery being used in tandem with an abnormal voltage threshold value being seen in the data; electrocautery use during this shoulder surgery is probable, and the likely cause of the premature end of a life.This will not be considered the root cause until further confirmation is received.No other relevant information has been received to date.
 
Event Description
After further review of data, it was determined that the battery life begins to rebound, indicating that a beginning of life impedance issue was the cause of the reported premature end of life.However, with this conclusion, it can be determined that no true premature end of life is occurring.This event is related to an increased duration of the high battery impedance experienced by cfx batteries during the beginning of life (bol) or an internal short that managed to ¿burn¿ itself out.This is not considered a fault of the device.No other relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMMETRY GEN MODEL 8103
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
dana sprague
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key15971171
MDR Text Key306810984
Report Number1644487-2022-01601
Device Sequence Number1
Product Code MUZ
UDI-Device Identifier05425025750504
UDI-Public05425025750504
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number8103
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received 11/18/2022
Initial Date FDA Received12/13/2022
Supplement Dates Manufacturer Received12/19/2022
01/23/2023
Supplement Dates FDA Received01/13/2023
02/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age33 YR
Patient SexMale
-
-