• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. PULSE GEN MODEL 106; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problems Premature Discharge of Battery (1057); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2022
Event Type  malfunction  
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanovas employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any defects or malfunctions.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
Surgeon reported he thought the patients generator was depleting prematurely.He felt like something could have been faulty with the generator.The patient had a generator replacement surgery.The suspect generator has been received by manufacturer and product analysis has not been completed to date.No other relevant information has been received to date.
 
Event Description
Product analysis (pa) for suspect generator was completed.Visual observations are most likely associated with manipulation of the device during the implant/explant procedures.An interrogation, a system diagnostic test, a voltage battery calculation, a comprehensive automated electrical evaluation, and a battery life calculation were performed.No anomalies were seen, and the device performed according to functional specifications.There were no performance, or any other type of adverse conditions found with the pulse generator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
dana sprague
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key15971613
MDR Text Key305665469
Report Number1644487-2022-01603
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750061
UDI-Public05425025750061
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/11/2023
Device Model Number106
Device Lot Number205624
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2022
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 11/18/2022
Initial Date FDA Received12/13/2022
Supplement Dates Manufacturer Received12/21/2022
Supplement Dates FDA Received01/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/25/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age36 YR
Patient SexMale
-
-