• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC ORBERA365¿ INTRAGASTRIC BALLOON SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC ORBERA365¿ INTRAGASTRIC BALLOON SYSTEM Back to Search Results
Model Number B-50012
Device Problems Deflation Problem (1149); Fluid/Blood Leak (1250); Mechanical Problem (1384); Unintended Deflation (4061)
Patient Problem Failure of Implant (1924)
Event Date 10/20/2022
Event Type  malfunction  
Event Description
Patient experienced greenish color urine.A endoscopic was performed and it was noted that the balloon was fully deflated.Balloon was removed successfully.Patient is doing fine.
 
Manufacturer Narrative
A review of the device labeling notes the following: the current orbera365¿ intragastric balloon system directions for use (dfu) addresses the known and anticipated potential events of "deflation and early removal "as follows: "the physiological response of the patient to the presence of the orbera365¿system balloon may vary depending upon the patient's general condition and the level and type of activity.The types and frequency of administration of drugs or diet supplements and the overall diet of the patient may also affect the response." "each patient must be monitored closely during the entire term of treatment in order to detect the development of possible complications.Each patient should be instructed regarding symptoms of deflation, gastrointestinal obstruction, ulceration and other complications which might occur, and should be advised to contact his/her physician immediately upon the onset of such symptoms." "complications: possible complications of the use of the orbera365¿ system include: -gastric discomfort, feelings of nausea and vomiting following balloon placement as the digestive system adjusts to the presence of the balloon.-patients reporting loss of satiety, increased hunger and/or weight gain should be examined endoscopically as this is indicative of an igb deflation.A patient who's deflated (i.E.Collapsed) igb has moved into the intestines must be monitored closely for an appropriate period of time (at least 2 weeks) to confirm its uneventful passage through the intestine.Abdominal or back pain, either steady or cyclic.Deflation and subsequent replacement." deflated device should be removed promptly.Possible adverse events: intestinal obstruction by the igb.An insufficiently filled igb or a leaking igb that has lost sufficient volume may be able to pass from the stomach into the small bowel.It may pass all the way into the colon and be passed with stool.However, if there is a narrow area in the bowel or adhesion formation, which may occur after previous surgery on the bowel, the igb may not pass and could cause a bowel obstruction.If this occurs, surgery or endoscopic removal could be required.Insufficient or no weight loss.Igb deflation (i.E.Collapse) and subsequent replacement.Additional information: the device has not been returned for analysis.A previous dhr review was completed for lot number, af04806 for (b)(4).The subject product met all specifications and requirements in effect at the time of manufacture.There was one other complaint in the apollo database against this lot number.
 
Manufacturer Narrative
Supplement # 1 medwatch submitted to the fda on 17/jan/2023.Additional information: a previous dhr review was completed for lot number, af04806 for (b)(4).The subject product met all specifications and requirements in effect at the time of manufacture.There was one other complaint in the apollo database against this lot number.Device evaluation summary: the device was returned to the apollo device analysis laboratory on 28/nov/2022.A deflated balloon with blue coloration on the shell was returned.Under microscopic analysis, the small hole on the shell has rounded edges and rolls inward.Due to the small hole, functional evaluation of the device could not be conducted.The complaint has been verified as it is uncertain how the small hole was created as the shell rolls inward which is not consistent with a puncture from a surgical instrument.Lab analysis was able to replicate the reported event of "deflation", as there is a small hole on the shell.The user effect of "deflation" is known and labeled possible adverse event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORBERA365¿ INTRAGASTRIC BALLOON SYSTEM
Type of Device
INTRAGASTRIC BALLOON SYSTEM
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC
1120 s. captail of texas hwy
bldg 1, ste 300
austin TX 78746
Manufacturer (Section G)
APOLLO ENDOSURGERY COSTA RICA, SRL
coyol free zone
building b 13.3
alajuela, cs CRI
CS   CRI
Manufacturer Contact
david hooper
1120 s. captail of texas hwy
bldg 1, ste 300
austin, TX 78746
5128523757
MDR Report Key15977592
MDR Text Key308532100
Report Number3006722112-2022-00117
Device Sequence Number1
Product Code LTI
UDI-Device Identifier10811955020725
UDI-Public(01)10811955020725(17)20231110(10)AF04806
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P140008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/10/2023
Device Model NumberB-50012
Device Catalogue NumberB-50012
Device Lot NumberAF04806
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/04/2022
Initial Date FDA Received12/14/2022
Supplement Dates Manufacturer Received11/04/2022
Supplement Dates FDA Received01/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/10/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age58 YR
Patient SexFemale
Patient Weight68 KG
-
-