There was an allegation of questionable elecsys rubella igg immunoassay results for 3 patient samples on a cobas e 801 analytical unit compared to a siemens assay.For patient 1, the elecsys rubella igg result was 17.2 iu/ml, interpreted as positive.The siemens assay result was negative.For patient 2, the elecsys rubella igg result was 91.9 iu/ml, interpreted as positive.The siemens assay result was negative.For patient 3, the elecsys rubella igg result was 27.2 iu/ml, interpreted as positive.The siemens assay result was negative.The initial elecsys assay results were reported outside of the laboratory.The e801 analyzer serial number is (b)(4).
|