• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH SLIDING MECHANISM; FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH SLIDING MECHANISM; FORCEPS Back to Search Results
Model Number 314.291
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Complainant part is not expected to be returned for manufacturer review/investigation.Occupation: initial reporter is a synthes employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Part # 314.291, lot # 14-5920, manufacturing site: (b)(4), release to warehouse date: 27 feb 2015, supplier: leitner ag, expiration date: n/a.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from japan reports an event as follows: it was reported that on an unknown date, the patient underwent the surgery with the products in question.During the operation check of the equipment, the connection between the main unit and various arms of the sliding reduction clamp in question could not be connected because the connection part of the main unit and the various arms did not fit.The diameter of the body side of the connection part and the diameter of the arm side did not seem to match in the first place.There are three types of arms, but all three have mismatched diameters and cannot be connected to the main unit.No further information is available.This report is for a sliding mechanism.This is report 1 of 4 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D9: device returned.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Part # 314.291, lot # 14-5920, manufacturing site: werk selzach logistik, release to warehouse date: 27 feb 2015, supplier: (b)(4).Expiration date: n/a.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that collinear reduction clamp sliding mechan was observed with signs of usage.A dimensional inspection for the collinear reduction clamp sliding mechan was not performed as it is not applicable to the complaint condition.A functional test was performed to device and the sliding mechanism of the clamp was found jammed.The complaint condition was able to be replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the collinear reduction clamp sliding mechan would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLIDING MECHANISM
Type of Device
FORCEPS
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
BACHLER FEINTECH AG
postfach
hölstein 4434
SZ   4434
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 
3035526892
MDR Report Key15988959
MDR Text Key308157839
Report Number8030965-2022-11172
Device Sequence Number1
Product Code HTD
UDI-Device Identifier10886982188779
UDI-Public(01)10886982188779
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number314.291
Device Catalogue Number314.291
Device Lot Number14-5920
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/25/2022
Initial Date FDA Received12/15/2022
Supplement Dates Manufacturer Received02/23/2023
03/30/2023
Supplement Dates FDA Received03/14/2023
04/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/27/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
HOHMANN-SHAPE ARM F/COLLIN REDUC CLAMP; PELVIC ARM F/COLLIN REDUC CLAMP; PERCUT-ARM F/COLLIN REDUC CLAMP
-
-