• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH REDUCT-FORCEPS-LRG W/POINT SOFTLO L222

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH REDUCT-FORCEPS-LRG W/POINT SOFTLO L222 Back to Search Results
Catalog Number 399.092
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/25/2022
Event Type  malfunction  
Event Description
Device report from synthes reports an event in portugal as follows: it was reported that the orthokit arrived at the local warehouse with a device broke.This report is for one (1) reduct-forceps-lrg w/point softlo l222 this is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Initial reporter occupation is a j&j employee.The device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part number: 399.092 lot number: t190297 manufacturing site: tuttlingen release to warehouse date: 04-apr-2020.A review of the device history records was performed for the finished device lot number and a and a nc was started due to a rough surface of the connection area between the two handles which led to less room for the instrument to move freely.The necessary actions to ensure the final product quality have been taken and documented in the appropriate quality system.The nc have no impact to the complaint condition.The final quality release criteria were met before this batch was released for distribution.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that reduct-forceps-lrg w/point softlo l222 has one of the jaws is broken.No other issues was found.A dimensional inspection was not performed for the aiming reduct-forceps-lrg w/point softlo l222 due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to excessive/unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the reduct-forceps-lrg w/point softlo l222 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed? yes, reviewed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDUCT-FORCEPS-LRG W/POINT SOFTLO L222
Type of Device
FORCEPS
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES TUTTLINGEN
unter hasslen 5
tuttlingen 78532
GM   78532
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15990604
MDR Text Key308115925
Report Number8030965-2022-11188
Device Sequence Number1
Product Code HTD
UDI-Device Identifier07611819220966
UDI-Public(01)07611819220966
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number399.092
Device Lot NumberT190297
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2022
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/26/2022
Initial Date FDA Received12/15/2022
Supplement Dates Manufacturer Received01/12/2023
Supplement Dates FDA Received02/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/04/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-