MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 24 FR.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 24 FR. Back to Search Results
Model Number A22041A
Device Problems Break (1069); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation where service confirmed the customer's complaint.During the device evaluation, a broken beak was found which was confirmed to be the cause of the reported issue.The investigation is still ongoing, a supplemental report will be submitted once the investigation is complete.Olympus will continue to monitor the field performance of this device.
 
Event Description
The customer reported to olympus that a resection sheath tip broke off in the patient¿s bladder during a bladder tumor removal.The tip was able to be retrieved and complete procedure.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer¿s investigation, the broken ceramic tip was due to damage of the insulation material of the sheath was caused by thermal mechanical overload, improper handling, mechanical impact like fall, shock or similar stress.Also noted were issues attributed to wear and tear.Please note that signs of fatigue or pre-damage, such as minute cracks, are often hard to spot.Based on the described damage pattern, it can be assumed the insulation tip's damage was caused by mechanical thermal influence.Unfortunately, it cannot be determined with certainty, whether there was previous damage on the device or any damage on the ceramic insulating insert was caused during last reprocessing or during last usage.The instruction for use (ifu) carries a warning that the ceramic tip can break due to mechanical loading or thermally induced straining.Thus, it is the responsibility of the user to inspect the instrument prior to every procedure.However, a definitive root cause could not be determined.The event can be detected/prevented by following the instructions for use which state: warning infection control risk properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing inspecting the product visually inspect the product.Make sure that it has: -- no corrosion -- no teeth marks -- no scratches ceramic insulation at distal end visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning risk of injury impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESECTION SHEATH, 24 FR.
Type of Device
RESECTION SHEATH
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15990673
MDR Text Key308110820
Report Number9610773-2022-00664
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761020961
UDI-Public04042761020961
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K931994
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22041A
Device Catalogue NumberA22041A
Device Lot Number20X01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/17/2022
Initial Date FDA Received12/15/2022
Supplement Dates Manufacturer Received12/22/2022
Supplement Dates FDA Received01/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-