Catalog Number CDS0705-NT |
Device Problem
Device Markings/Labelling Problem (2911)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/22/2022 |
Event Type
malfunction
|
Event Description
|
This is filed to report a labeling problem.It was reported the labeling on the brown shipping box indicated the product was an nt clip.However, after the box was opened, it was observed the device inside was an xt clip.No additional information was provided.
|
|
Manufacturer Narrative
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Manufacturer Narrative
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents reported from this lot.The investigation determined that there was a mix-up of brown corrugated shipper boxes when the product was exchanged between 2 distributors in brazil.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
|
|