MAUDE Adverse Event Report: ABBOTT AUTOMATION SOLUTIONS GMBH GLP LOADER MODULE; CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT AUTOMATION SOLUTIONS GMBH GLP LOADER MODULE; CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE Back to Search Results
Catalog Number 06T12-01
Device Problem Product Quality Problem (1506)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2022
Event Type  malfunction  
Manufacturer Narrative
During investigation it was found that the reason the 16 mm sample tubes were getting stuck in the glp disposal tube is because the disposal tube was found to be deformed and not quite round.The deformed disposal tube lead to sample tubes getting stuck causing back-up and potential spillage of the sample tubes and its contents.The cause for the deformed tube was determined to be related to the material of the disposal tubing (pvc).The material was found to be too flexible.Because the tubing is deformed and flexible, they become bent when being inserted, resulting in less space for the sample tubes to slide into the waste bin.As an interim action the disposal tubes will be checked at the supplier via a test procedure for incoming inspection and for in-process controls.As a long-term solution an alternate tubing material will be found that cannot get easily bent.Based on the investigation, a deficiency was identified with the glp disposal tube (list number 06t24-01) for the glp loader module (lm), serial (b)(4).This report is being filed on an international product, glp loader module, list number 06t12-01, which has a same/similar component of the modular glp systems track registered in the us, list number 04z96-51 note: the glp systems track is not yet distributed in the us.
 
Event Description
The customer reported samples were stuck in the waste chute of the glp loader module.The customer stated the whole waste chute was removed and the glp disposal tube was emptied.No further issues were reported.There was no patient involvement and no reported impact to patient management or user safety.
 
Event Description
The customer reported samples were stuck in the waste chute of the glp loader module.The customer stated the whole waste chute was removed and the glp disposal tube was emptied.No further issues were reported.There was no patient involvement and no reported impact to patient management or user safety.
 
Manufacturer Narrative
The long-term solution is to remove the tubing to utilize the full diameter of the disposal pathway, which will reduce the likelihood of tubes getting caught in the disposal pathway.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLP LOADER MODULE
Type of Device
CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE
Manufacturer (Section D)
ABBOTT AUTOMATION SOLUTIONS GMBH
sachsenkamp 5
hamburg 20097
GM  20097
Manufacturer (Section G)
ABBOTT AUTOMATION SOLUTIONS GMBH
sachsenkamp 5
hamburg 20097
GM   20097
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key15992901
MDR Text Key308074799
Report Number3023268435-2022-00024
Device Sequence Number1
Product Code JQP
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K213486
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number06T12-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/30/2022
Initial Date FDA Received12/15/2022
Supplement Dates Manufacturer Received08/28/2023
Supplement Dates FDA Received09/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
GLP DISPOSAL TUBE, 06T24-01, UNKNOWN; GLP DISPOSAL TUBE, 06T24-01, UNKNOWN
-
-