Model Number 16-02-80
Device Problem Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2022
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(4).Livanova through follow-up communication with the customer, learn that additional sockets has been tested including sockets with test and reset functions and all of them caused the tripping of the mains circuit of the hospital.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The unit would power up as usual and when the temperature was lowered it would trip the power.The failure was traced back to a short circuit of the cooling compressor due to loss of refrigerant gas.The unit was removed from service and will be scrapped.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t device tripped the mains circuit of the hospital during disinfection and cleaning processes.There was no patient involvement.
 
Manufacturer Narrative
The reported event is a known issue: breach of the cooling coil can be generated due to stirrer flow in the tank.Consequently, the water will enter the cooling circuit and the compressor unit.This potential failure mode will cause immediate damage of the cooling compressor system.The compressor failure leads to an increase of the leakage current of the device and the circuit breaker will trip.Complaints database analysis revealed no similar case notified since unit installation in 2017.Sap analysis revealed that on august 22nd, 2019 it was performed the erosion kit upgrade of the devices that includes the new design of the pump head of the stirrer motor to prevent a water flow directed to a narrow area of the evaporator coil.The issue was claimed about more than three years after the upgrade, and, most probably, the condition of the evaporator was already compromised before the upgrade to such an extent that the copper was rather degraded.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See intial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80004
GM   80004
Manufacturer Contact
enrico greco
14401 w 65th way
arvada, CO 80004
MDR Report Key15994376
MDR Text Key305843219
Report Number9611109-2022-00679
Device Sequence Number1
Product Code DWC
UDI-Device Identifier04033817901099
UDI-Public(01)04033817901099(11)170320
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/22/2022
Initial Date FDA Received12/16/2022
Supplement Dates Manufacturer Received03/04/2023
Supplement Dates FDA Received03/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/20/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-