BIOMET UK LTD. OXF CMNTLS IMPLANT INSERT; KNEE INSTRUMENT Back to Search Results
Model Number N/A
Device Problems Fracture (1260); Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2022
Event Type  malfunction  
Event Description
It was reported that during surgery, while impacting the tibial implant, a piece of the instrument broke off.Nothing left in the patient.No consequences or impact to the patient.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Manufacturer Narrative
(b)(4).Report source ¿ foreign ¿ canada.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.Visual examination of the returned product confirms the adjustable back foot has fractured off.The instrument exhibits indentations on the top of the main body.There is also evidence of indentations on the back of the instrument around the adjusting area.The instrument has been in the field for approximately 13 years and 3 months.Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).The product was not returned.The instrument has potentially been in the field for approximately 13 years and 1 month.Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF CMNTLS IMPLANT INSERT
Type of Device
KNEE INSTRUMENT
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key15994423
MDR Text Key308089577
Report Number3002806535-2022-00504
Device Sequence Number1
Product Code HRY
UDI-Device Identifier05019279305223
UDI-Public(01)05019279305223(11)170602(10)ZB090713
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number32-422097
Device Lot NumberZB090713
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/17/2022
Initial Date FDA Received12/16/2022
Supplement Dates Manufacturer Received12/21/2022
02/22/2023
Supplement Dates FDA Received01/04/2023
03/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-