ZIMMER SWITZERLAND MANUFACTURING GMBH STEM STELCOR 4 R; ANATOMIC PRIMARY STEM Back to Search Results
Model Number N/A
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problem Failure of Implant (1924)
Event Date 11/15/2004
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: artek cup me 38/54; item# :91.36.38-54; lot#: 2049338; metasul-femoral head 38 m; item#: 19.38.06; lot#: 2065925.Report source: foreign: germany.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that the patient underwent an initial right total hip arthroplasty.Approximately 3 years later a revision surgery was performed due to loosening.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h1, h2, h3, h6, h10.No product was returned or pictures provided: product evaluation could not be performed.Review of the manufacturing records identified no deviations or anomalies during manufacturing related to the reported event.Device used for treatment.Medical records were received and reviewed.The patient is male, at the time of the reported event he is 44 years old.The patient underwent a first bilateral tha because of dysplastic arthrosis at both sides.Both cups have been implanted with a low inclination (as well as a low anteversion.With the available data a definitive root cause cannot be established.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM STELCOR 4 R
Type of Device
ANATOMIC PRIMARY STEM
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key15995032
MDR Text Key305590991
Report Number0009613350-2022-00658
Device Sequence Number1
Product Code MRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/15/2003
Device Model NumberN/A
Device Catalogue Number360029040
Device Lot NumberB273058
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/23/2022
Initial Date FDA Received12/16/2022
Supplement Dates Manufacturer Received03/14/2023
Supplement Dates FDA Received03/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/31/1999
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexMale
-
-