• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS VORTEX PORT; VORTEX LP DUAL TITANIUM PORT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS VORTEX PORT; VORTEX LP DUAL TITANIUM PORT SYSTEM Back to Search Results
Model Number LVTX5213
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Air Embolism (1697)
Event Date 11/28/2022
Event Type  Injury  
Manufacturer Narrative
It was reported that the disposable device is not available to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event Description
An  ir tech manager reported an issue with a lp titanium vtx 12 fr det 12f intro port.During a placement procedure, the patient experienced an air embolism, upon removal of the wire and dilator, prior to insertion of the catheter.The patient was stabilized and the air embolism resolved by the time the interventionist intervened.The procedure was completed.After the procedure, the provider had determined after the embolism occurred that they should have used a valved sheath, and not the non-valved.
 
Manufacturer Narrative
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.No sheath accessory device was returned to angiodynamics for evaluation since there was no report of device malfunction during use.The customer's reported complaint description of air bubbles/embolism were introduced into the patient's vasculature cannot be confirmed given the patient centric nature of this adverse event.No sheath/dilator accessory device was returned for evaluation since there was no report of device malfunction during the procedure.Based on the event description details the likely root cause of the air embolism is end user did not sufficiently place thumb over the exposed oriffice of the sheath hub after the dilator/guidewire were removed and before the catheter tubing could be inserted.This is cautioned against in the dfu.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.  the review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the instructions for use, which is supplied to the user with this item number, contains the following statements.Catheter placement considerations: unclamp guidewire and advance dilator/sheath over the exposed "j" wire.Withdraw vessel dilator and "j" wire, leaving sheath in place.Caution: to prevent air embolism, place thumb over exposed orifice of sheath.Warning: for chest placement, avoid medial catheter placement.Instructions for implantation of the vortex® mp port into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route to the subclavian vein.Potential complications.Use of angiodynamics port systems involve potential risks normally associated with the insertion or use of any implanted device or indwelling catheter including but not limited to: infection; occlusion; thrombophlebitis, pneumothorax; catheter malposition; migration and inadequate anchoring; hemorrhage; vessel trauma, including puncture, laceration, and erosion of vessel and the skin; catheter pinch-off (compression of the catheter between the clavicle and the first rib); hematoma; clot formation; catheter fragmentation; embolization; cardiac arrhythmia; cardiac puncture; cardiac tamponade; fibrin sheath, endocarditis; implant rejection; thoracic duct injury; thromboembolism; peritonitis; thrombosis; and drug extravasation (leakage).A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VORTEX PORT
Type of Device
VORTEX LP DUAL TITANIUM PORT SYSTEM
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587805
MDR Report Key15995993
MDR Text Key305604867
Report Number1317056-2022-00145
Device Sequence Number1
Product Code LJT
UDI-Device IdentifierH787LVTX52130
UDI-PublicH787LVTX52130
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K010767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberLVTX5213
Device Catalogue NumberLVTX5213
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/28/2022
Initial Date FDA Received12/16/2022
Supplement Dates Manufacturer Received02/21/2023
Supplement Dates FDA Received02/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-