• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH STEM STELCOR 4 L; HIP PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH STEM STELCOR 4 L; HIP PROSTHESIS Back to Search Results
Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Metal Related Pathology (4530)
Event Date 09/02/2020
Event Type  Injury  
Event Description
It was reported that the patient underwent a bilateral total hip arthroplasty and approximately 21 years later, during medical checkups, elevated metal ions (cobalt and chromium) were identified in blood work.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Manufacturer Narrative
(b)(4).Concomitant medical products: artek cup me 38/52; item#: 91.36.38-52; lot#: 2057482; metasul-femoral head 38 m; item#: 19.38.06; lot#: 2065925.Report source: foreign: germany.Multiple mdr reports were filed for this event, please see associated reports: 0009613350 - 2022 - 00665, 0009613350 - 2022 - 00666.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b3, b4, b5, g3, g6, h1, h2, h6, h10.No product was returned or pictures provided: product evaluation could not be performed.Review of the manufacturing records identified no deviations or anomalies during manufacturing related to the reported event.Devices used for treatment.Medical records were received and reviewed.The patient is male, at the time of the reported event he is 44 years old.The patient underwent a first bilateral tha because of dysplastic arthrosis at both sides.Both cups have been implanted with a low inclination as well as a low anteversion.With the available data a definitive root cause cannot be established.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM STELCOR 4 L
Type of Device
HIP PROSTHESIS
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key16003446
MDR Text Key305684965
Report Number0009613350-2022-00664
Device Sequence Number1
Product Code MRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/15/2003
Device Model NumberN/A
Device Catalogue Number36.00.19-040
Device Lot NumberB273055
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/23/2022
Initial Date FDA Received12/19/2022
Supplement Dates Manufacturer Received03/14/2023
Supplement Dates FDA Received03/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/31/1999
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient Outcome(s) Other;
Patient SexMale
-
-