Updated fields: b4, d9, e1(site country), g3, g6, h2, h3, h4, h6 ( type of investigation, investigation findings, component codes , investigation conclusion), h10, h11.Corrected fields: h6 (health effect ¿ clinical code, health effect ¿ impact codes).Additional information: event site postal code: 12120, email: skornthip@yahoo.Co.Th.A getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue.To fix the issue, the fse replaced front end module pcb, and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released and cleared for clinical service.
|