DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE G PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-52
Device Problems Component Incompatible (1108); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2022
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.Due to character restrictions telephone number :(b)(6).
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) , the cardiosave intra-aortic balloon pump (iabp) unit was found to have a fill manifold failure.There was no patient involvement.
 
Manufacturer Narrative
The fse that encountered the issue determined that the unit had a faulty k3 solenoid on the fill manifold after failing the "all manifold tests".The fse replaced the helium fill manifold assemble (0104-00-0014).The fse performed the all manifold tests.All tests passed.The iabp was returned to the customer and cleared for clinical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE G PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16003962
MDR Text Key308192087
Report Number2249723-2022-03197
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-52
Device Catalogue Number0998-00-0800-52
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/01/2022
Initial Date FDA Received12/19/2022
Supplement Dates Manufacturer Received05/23/2023
Supplement Dates FDA Received05/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-