Model Number 0998-00-0800-52 |
Device Problems
Component Incompatible (1108); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.Due to character restrictions telephone number :(b)(6).
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) , the cardiosave intra-aortic balloon pump (iabp) unit was found to have a fill manifold failure.There was no patient involvement.
|
|
Manufacturer Narrative
|
The fse that encountered the issue determined that the unit had a faulty k3 solenoid on the fill manifold after failing the "all manifold tests".The fse replaced the helium fill manifold assemble (0104-00-0014).The fse performed the all manifold tests.All tests passed.The iabp was returned to the customer and cleared for clinical use.
|
|