• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INT'L LTD. 8010379 DXTEND METAGLENE; DELTA XTEND IMPLANTS : SHOULDER METAGLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INT'L LTD. 8010379 DXTEND METAGLENE; DELTA XTEND IMPLANTS : SHOULDER METAGLENE Back to Search Results
Model Number 130760000
Device Problem Fracture (1260)
Patient Problem Insufficient Information (4580)
Event Date 11/07/2022
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
This complaint is for a blind unit.It was found during the visual examination performed on (b)(6) 2022 that there was a small chuck of the ha coating missing from the metagene.Follow-up was conducted to determine when the ha coating fell off, but there was no additional information available regarding the device and/or procedure.
 
Manufacturer Narrative
Product complaint # : (b)(4).Investigation summary : the device associated with this report was returned for analysis.Visual analysis of the device found that a small portion of ha coating was chipped from the device., confirming the reported allegation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.Based on the inability to find any nc¿s against the provided product code/lot code combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND METAGLENE
Type of Device
DELTA XTEND IMPLANTS : SHOULDER METAGLENE
Manufacturer (Section D)
DEPUY INT'L LTD. 8010379
st. anthony's rd
leeds LS11 8 DT
UK  LS11 8 DT
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st. anthony's rd
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16005699
MDR Text Key306891079
Report Number1818910-2022-25730
Device Sequence Number1
Product Code HSD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192855
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number130760000
Device Catalogue Number130760000
Device Lot Number3794684
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/06/2022
Initial Date FDA Received12/19/2022
Supplement Dates Manufacturer Received01/05/2023
01/20/2023
Supplement Dates FDA Received01/09/2023
01/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexFemale
-
-