RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Charging Problem (2892)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/22/2022 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint from the customer on the v60 indicating that the device would not operate on ac power but would run on battery power.It was reported that there was no patient involvement at the time the issue was discovered.The remote service engineer (rse) troubleshot the device per the service guide and determined that the device had 24vdc coming from the power supply.The rse advised that the device may have a faulty pm pcb.An onsite work order was created.The investigation is ongoing.
|
|
Manufacturer Narrative
|
Multiple attempts to retrieve device repair, and operational status has yielded no response from the customer. the complaint will be processed for closure.If additional information becomes available at a later date, the complaint will be reopened, and a supplemental report will be submitted.
|
|