Catalog Number 09180443001 |
Device Problem
Application Program Problem (2880)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/23/2022 |
Event Type
malfunction
|
Event Description
|
There was an allegation of a software issue for the cobas infinity core software, version 3.03.11.There was reportedly a delay with rule engine execution of ten seconds to 1.5 hours which causes incorrect results to be validated and released to the laboratory information system (lis).It has been confirmed that no patient was harmed because of this issue.
|
|
Manufacturer Narrative
|
The investigation verified the allegation as a software issue in cobas infinity, as the rule engine process is taking too long to process the events which caused delays in the rule execution.This delay allows the automatic validation process to run before the rule is executed.
|
|
Manufacturer Narrative
|
The investigation determined that the causal factor of the reported allegation is the low performance of the rule engine module.
|
|
Search Alerts/Recalls
|