Catalog Number 955626
Device Problem Overheating of Device (1437)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  Injury  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that "overheating" of a prismax machine was observed after being used for extended periods.In all instances, extracorporeal blood was manually returned to the patient.Blood clots were observed, and the patients were transfused.No additional information is available.
 
Manufacturer Narrative
Additional information added to h3, h6 and h10.The actual device was evaluated on site by a biomedical technician.No overheating of the machines was confirmed, and no device malfunction was identified.The reported condition was not verified.The cause of the condition could not be determined; however, the most likely the cause was due to a ¿lack of nurse education¿.The unit where the issue occurred also "reflected it was nursing error".A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAX MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - BROOKLYN PARK
7601 northland drive
n suite 170
minneapolis MN 55428
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16014344
MDR Text Key305803753
Report Number3003504604-2022-00038
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190910
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number955626
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/21/2022
Initial Date FDA Received12/20/2022
Supplement Dates Manufacturer Received01/20/2023
Supplement Dates FDA Received01/20/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-