MAUDE Adverse Event Report: MEDTRONIC XOMED INC. NUVENT¿; INSTRUMENT, ENT MANUAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. NUVENT¿; INSTRUMENT, ENT MANUAL SURGICAL Back to Search Results
Model Number 1830717SPH
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2022
Event Type  malfunction  
Manufacturer Narrative
No parts have been returned to medtronic for analysis.Codes b17, c20, d15 are applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that there was issues with several trackable instruments being recognized by the system.When plugged in, the site got a green indication that the system read the instrument.However, tracking details did not show the instrument, and the patient tracker was between.5 and.7.There was no patient involvement.No, troubleshooting / manual workaround resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVENT¿
Type of Device
INSTRUMENT, ENT MANUAL SURGICAL
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
glen belmer
6743 southpoint drive north
jacksonville, FL 32216
6122713209
MDR Report Key16017972
MDR Text Key308248538
Report Number1045254-2022-00825
Device Sequence Number1
Product Code LRC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132297
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1830717SPH
Device Catalogue Number1830717SPH
Device Lot Number0225322939
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/09/2022
Initial Date FDA Received12/20/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/19/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-