• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-E
Device Problem Gas Output Problem (1266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2022
Event Type  malfunction  
Event Description
The customer reported to olympus that gas was leaking from the high flow insufflation unit.The issue was found during maintenance.There were no reports of patient involvement.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found leakage from the first regulator and leakage from the connector unit.A review of the device history record (dhr) found no deviations that could have caused or contributed to the reported device problem.The dhr confirmed that the subject device was shipped in accordance with the specifications.A definitive root cause for this issue was not established.However, it is probable that the issue occurred because of faulty first regulator unit and faulty connector.The cause of the fault was unknown.Olympus will continue to monitor field performance for this device.A supplemental report will be submitted if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16018324
MDR Text Key308176766
Report Number3002808148-2022-05434
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/22/2022
Initial Date FDA Received12/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/23/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-