Model Number CYF-VHA
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus, there was an e311 error (the white balance has not been set).During the inspection and testing of the returned device, there was a foreign object in the illumination lens.There were no reports of patient harm associated with this event.This medical device report (mdr) is being submitted to capture the reportable malfunction found during the device evaluation.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer¿s allegation was not confirmed.In addition to the findings reported in event, scratches were found throughout the device, wrinkles were found in the insertion tube, the bending angle was insufficient due to elongation of the angle wire, the light guide bundle was burning, there was play on the curved pipe and the universal cord was collapsed.There was yellowing observed and the image was dark to due the burning of the light guide bundle.Additional information has been requested regarding this event.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or when additional information becomes available.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and from the burn-like state of the illumination lens, it is likely that the foreign matter could not be removed even after proper reprocessing due to physical damage to the illumination lens.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16019625
MDR Text Key308235918
Report Number3002808148-2022-05445
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-VHA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/20/2022
Initial Date FDA Received12/20/2022
Supplement Dates Manufacturer Received12/22/2022
Supplement Dates FDA Received01/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-