• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS INC. IMPRESS®.BRAIDED CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS INC. IMPRESS®.BRAIDED CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number 59038UAC
Device Problems Fracture (1260); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2022
Event Type  Injury  
Event Description
The account alleges that during a uterine fibroid embolization procedure.The tip of the catheter broke while withdrawing it from the patient.A snare was used to successfully retrieve the broken pieces.No additional patient consequence to report.
 
Manufacturer Narrative
The suspect device is not expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Manufacturer Narrative
The suspect medical device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and 1 similar complaints for this lot number was identified.Should the device be returned later, the investigation will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS®.BRAIDED CATHETER
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS INC.
14646 kirby drive
houston TX 77047
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS INC.
14646 kirby drive
houston TX 77047
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
MDR Report Key16019733
MDR Text Key305871223
Report Number3010665433-2022-00127
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K053171
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/08/2023
Device Catalogue Number59038UAC
Device Lot NumberE2004961
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/08/2022
Initial Date FDA Received12/20/2022
Supplement Dates Manufacturer Received02/13/2023
Supplement Dates FDA Received02/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/13/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-