Reports are being submitted on the oer-4 aer and the scope.Please refer to the following reports: patient identifier of (b)(6) is related to model number: oer-4, serial number: (b)(4).Patient identifier of (b)(6) is related to model number: gif-h290z, serial number: (b)(4).This event is under investigation.A supplemental report will be submitted upon receiving additional information.
|
The customer reported that the main board of the oer-4 automatic endoscope reprocessor (aer) was recently replaced; however, after the replacement, the maintenance tool could not be connected to the aer.The issue was resolved after reinstalling the oer-4 program.When reviewing the work, the customer found that a board for the oer-3 was used.Between (b)(6) 2022 and (b)(6) 2022, the aer was used with the oer-3 board.No errors occurred and it was unclear whether the aer worked properly and cleaned in the correct process.Multiple scopes, including the evis lucera elite gastrointestinal videoscope, were reprocessed using the aer during this time.There was no reported patient harm or impact due to this event.
|