There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in melbourne, australia.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue: patient pump 2 did not start, except with a finger push.In order to fix the issue, the patient pump 2 was replaced with a new one.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|