• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC SHILLA GROWTH GUIDANCE SYSTEM; GROWING ROD SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC SHILLA GROWTH GUIDANCE SYSTEM; GROWING ROD SYSTEM Back to Search Results
Model Number 7676530
Device Problems Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Nervous System Problem (4426)
Event Date 11/22/2022
Event Type  Injury  
Event Description
Information was received from a healthcare provider (hcp) via manufacturer representative regarding spinal product used in spinal fi xation therapy for premature scoliosis.It was reported that posterior shilla method was performed after dissection of t7/8/9/10 by previous method.After correction, the left lower extremity monitoring completely disappeared.After removal of the left and right rods, it was checked using the o-arm and a puncture of the left t8ps was highly suspected, so it was removed.Moreover, the surgical (hemostatic agent) used during the posterior release was removed because it was corrected while it was still placed in the epidural.On the day of the event, correction was abandoned, and it ended up with unilateral rod fixation of the left t7-10.There was a complete loss of the left lower extremity on spinal monitoring and no further complications or symptoms were reported.
 
Manufacturer Narrative
Change in surgical procedure due to product malfunction.Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received via manufacturer representative that there is no damage to the screw.The trouble was caused by the procedure, not by the product.
 
Manufacturer Narrative
Review of this mdr and/or additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event did not and does not meet the reporting requirements stipulated in 21 cfr 803.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILLA GROWTH GUIDANCE SYSTEM
Type of Device
GROWING ROD SYSTEM
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key16021511
MDR Text Key305874079
Report Number1030489-2022-01122
Device Sequence Number1
Product Code PGM
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K140750
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7676530
Device Catalogue Number7676530
Device Lot NumberCA16F026
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/22/2022
Initial Date FDA Received12/21/2022
Supplement Dates Manufacturer Received12/22/2022
Supplement Dates FDA Received01/31/2023
Date Device Manufactured02/16/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-