Model Number ALLURA FC
Device Problem Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2022
Event Type  malfunction  
Event Description
It has been reported to philips that the software was corrupted - failed to start.The issue was found during clinical use.No harm has been reported to philips.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.The philips field service engineer (fse) inspected the system onsite confirmed that the system failed to start due to corrupted software.The fse reloaded the latest software.After reloading the software, system was returned to use in good working order.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA FC
Type of Device
SYSTEM, X-RAY, FLUOROSCOPIC, IMAGE-INTENSIFIED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16022120
MDR Text Key306559985
Report Number3010685285-2022-00007
Device Sequence Number1
Product Code JAA
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K022899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA FC
Device Catalogue Number723003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2022
Initial Date FDA Received12/21/2022
Supplement Dates Manufacturer Received04/24/2023
Supplement Dates FDA Received09/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/10/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-