• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION AK 98 MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION AK 98 MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955607
Device Problems Improper or Incorrect Procedure or Method (2017); Improper Flow or Infusion (2954)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2022
Event Type  malfunction  
Event Description
It was reported that during treatment with an ak 98 machine, the operator accidentally programmed the heparin delivery for a rate of 5 ml for hour, believing that the rate set was 5 u.I.Per hour.This led to an excessive heparin delivery to patient.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
The device was not received for evaluation; however, the event description clearly identify that the nurse set the heparin parameter believing that the unit of measurement was iu and not ml, causing an excessive heparin administration.Based on analysis of the ak98 operator's manual can be ascertained that the manual and the machine reports clearly that the measurement unit used to measure the heparin is ml and iu.There was no malfunction of the ak98 device and the machine worked according to specifications.The cause of the reported condition was associated with a use error when setting the heparin parameter.A service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AK 98 MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEDOLLA MODENA MONITORS
via modenese, 66
medolla modena 41036
IT   41036
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16023117
MDR Text Key308175302
Report Number9616240-2022-00026
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201809
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955607
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/01/2022
Initial Date FDA Received12/21/2022
Supplement Dates Manufacturer Received01/20/2023
Supplement Dates FDA Received01/26/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-