Model Number 3100 A |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/30/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Vyaire medical file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
It was reported to vyaire medical that piston displacement and pieces of bellows in circuit occurred on the 3100 a device.The customer reported there was no patient involvement associated with the reported event.
|
|
Manufacturer Narrative
|
Results of investigation: vyaire medical field service technician went onsite to evaluate the device.Patient removed circuit from unit before giving to biomed.Placed known good circuit on.Bypassed humidifier & blender.Ran patient circuit check.Ran performance check.Both passed.Dynamic displacement indicator positioning in specification with rubber stopper in & removed.Re-ran performance check and allowed vent to run on driver.No alarms present.Mean airway pressure & amplitude stable.All tests passed required criteria.Unit meets factory specifications.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|