• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE, 30°, 4 MM; HYSTERORESECTOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE, 30°, 4 MM; HYSTERORESECTOSCOPE Back to Search Results
Model Number A22002A
Device Problem Melted (1385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus for evaluation.During inspection and testing, the allegation of cloudy image was not confirmed.In addition to the finds reported, the insertion section outer tube was bent.Additional information has been requested regarding this event.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if additional information becomes available.
 
Event Description
An olympus employee reported on behalf of the customer, the telescope displayed a cloudy image.There was no report of user or patient harm associated with this event.During incoming inspection, the light guide connector burnt light guide connector.This medwatch is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
This medwatch is being supplemented with additional information obtained and the manufacturer's final investigation results.During the inspection, the reported event was confirmed.In addition, other defects were identified during the device inspection: blurry image, light guide connector was found burnt and the outer tube was bent.The reported failure mode/identified defects can be attributed to incorrect reprocessing.Remaining residues have burnt in due to the light energy and the associated heating.Bent outer tube can be attributed to the user error, improper handling and application of excessive force.A manufacturing and quality control review was performed for the affected serial number without showing any non-conformities or deviations regarding the described issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 30°, 4 MM
Type of Device
HYSTERORESECTOSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16030492
MDR Text Key308537197
Report Number9610773-2022-00713
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K897003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22002A
Device Catalogue NumberA22002A
Device Lot Number675276
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/24/2022
Initial Date FDA Received12/21/2022
Supplement Dates Manufacturer Received03/20/2023
Supplement Dates FDA Received04/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/07/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-