Manufacturer Narrative
Investigation: per the customer, the patient drinks an 8 oz glass of oj every morning including the day of the collection.A review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.Investigation is in process.A follow up report will be provided.
 
Event Description
The customer reported that they had to end the continuous mononuclear cell (cmnc) collection procedure after 24 minutes because they needed to take the patient for a ekg due to possible citrate toxicity.The citrate toxicity was resolved by increasing calcium iv infusion.The procedure was not continued based on the results of the ekg.The patient was admitted for observation and discharged.The customer declined to provide patient identifier.The collection set is not available because it was discarded by the customer.
 
Manufacturer Narrative
Investigation: per the customer, the patient drinks an 8 oz glass of oj every morning including the day of the collection.A review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.A disposable complaint history search was performed for this lot and found no reports for similar issues on this lot.According to therapeutic apheresis: a physician's handbook, adverse events occur during therapeutic procedures with a frequency of 4.8%.Some of the most common reactions include fever, urticaria, hypocalcemic symptoms, pruritus, dyspnea, tachycardia, and mild hypotension transient hypocalcemia associated with apheresis is usually well tolerated.Symptoms often show as paresthesia (tingling) but patients may also experience unusual taste, nausea, lightheadedness, shivering, and tremors.Severe hypocalcemia may also cause muscle contractions and can progress to tetany and seizures if hypocalcemia escalates and is not corrected.Root cause: a root cause assessment was performed for the reported citrate reactions.These reactions occur due to decreased ionized calcium in circulation as a result of exogenous citrate administered during the apheresis procedure and are influenced by donor physiology, the rate of ac infusion, and/or the length of the procedure.These symptoms may be treated with oral or intravenous calcium supplements or by adjusting the ac infusion rate.
 
Event Description
The customer reported that they had to end the continuous mononuclear cell (cmnc) collection procedure after 24 minutes because they needed to take the patient for a ekg due to possible citrate toxicity.The citrate toxicity was resolved by increasing calcium iv infusion.The procedure was not continued based on the results of the ekg.The patient was admitted for observation and discharged.The customer declined to provide patient identifier.The collection set is not available because it was discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRA OPTIA
Type of Device
SPECTRA OPTIA EXCHANGE SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key16035380
MDR Text Key306039970
Report Number1722028-2022-00422
Device Sequence Number1
Product Code GKT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK150251
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number12220
Device Lot Number2207223130
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/01/2022
Initial Date FDA Received12/22/2022
Supplement Dates Manufacturer Received02/10/2023
Supplement Dates FDA Received02/10/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age74 YR
Patient SexFemale
Patient Weight87 KG
-
-