SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problems Inflation Problem (1310); Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus, the high flow insufflation unit was slow to inflate, does not hold pressure or was unstable.There were no reports of patient harm associated with this event.
 
Manufacturer Narrative
The device has not been received to date.Additional information has been requested regarding this event.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and the device evaluation.Please see updates to d9, h3, h4, h6 and h10.The device was returned to olympus for inspection, and the customer's complaint was not confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, probable causes of the insufflator being slow to inflate and not hold pressure include, a faulty main board or connection failure of the connecting tube connected to the tube pressure sensor.However, the reported event was not duplicated during the device repair.Therefore, a final root cause was unable to be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16039716
MDR Text Key308466210
Report Number3002808148-2022-05535
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/23/2022
Initial Date FDA Received12/22/2022
Supplement Dates Manufacturer Received01/31/2023
Supplement Dates FDA Received02/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/28/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-