The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction pressure dome leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot l004 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot l004 shows no trends.Trends were reviewed for complaint category, pressure dome leak.No trends were detected for this complaint category.This assessment is based on the information available at the time of the investigation.At the time of this report, the analysis of the returned kit and smart card is still in progress.A supplemental report will be filed when the analysis is complete (b)(4).(b)(6) 2022.
|
The complaint kit was returned for investigation.The smart card was not returned.A visual inspection of the kit found no obvious manufacturing issues.Inspection of the collect, return, and system pressure domes did not identify a blood leak.The diaphragms on the collect, return, and system pressure domes were properly seated and exhibited no damage.In addition, the pressure domes were not damaged.The pressure domes were installed onto a pressure sensor to assess fit, all pressure domes latched into the circumferential groove around the sensor as expected.The three pressure domes were pressure tested to check for leaks and no leaks were identified.The customer complaint was not verified.No manufacturing related defects were identified during this investigation.No further action is required at this time.This investigation is now complete.Comp-037692 h.M.22 mar 2023.
|