(b)(4).Device evaluation by manufacturer: the aquabeam console was returned for investigation of the reported event.Functional testing was able to reproduce the reported "e03 - console error" upon priming.The error prevented the system to initialize an aquablation simulation.The current user manual, um0101-00 rev.F aquabeam robotic system user manual, states: table 5 system detected errors and faults "e03 - console error release foot pedal and click x.If error persists, turn off and turn on console." a review of the device history record (dhr) for aquabeam robotic system/serial number (b)(4) and the aquabeam console / lot number 22c01992 was performed, which confirmed that there were no nonconformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system and its associated component met all design and manufacturing specifications when released for distribution.The root cause of the reported event has been determined to be due to manufacturing-related issues associated with a third-party supplier.A corrective action and preventive action has been initiated by procept to determine appropriate actions.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|