AIZU OLYMPUS CO., LTD. CHOLEDOCHO VIDEOSCOPE; CHOLEDOCO VIDEOSCOPE Back to Search Results
Model Number CHF-V2
Device Problem Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned for investigation.Upon evaluation of the device, the reported issue was confirmed.In addition, curved rubber sagging, curved rubber adhesive missing, continuity failure, flexible tube of the insertion section crushed, wrinkles in the insertion tube due to external factors, forceps failure to insert, channel cleaning brush failure to be inserted, poor insulation performance at the tip, and forceps plug cap scrape were observed.Lastly, scratches were found on multiple device components.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
A user facility reported to olympus that a choledocho videoscope tip rubber deterioration, peeling, and cutting were noted.During a standard service inspection of the customer returned device, the forceps plug mouthpiece was scrapped.This report is to capture the reportable malfunction found at inspection.There was no patient harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the forceps plug mouthpiece being scratched was stress caused by of repeated use, external factors, or handling.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHOLEDOCHO VIDEOSCOPE
Type of Device
CHOLEDOCO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16046081
MDR Text Key308357275
Report Number9610595-2022-06063
Device Sequence Number1
Product Code FBN
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCHF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/01/2022
Initial Date FDA Received12/25/2022
Supplement Dates Manufacturer Received01/13/2023
Supplement Dates FDA Received02/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-